Determination of fluid leakages in the different screw-retained implant-abutment connections in a mechanical artificial mouth
Autor/a
Data de publicació
2015-07-15ISSN
0957-4530
Resum
This study shows the potential risk of microfiltration between two different types of implant-abutment connections screwed at 45 Ncm: external and internal. For the first time the use of a mechanical artificial mouth is used with the values (compression and torsion loads with a frequency of 2 Hz) of the human chewing. The mechanical tests were performed with an artificial saliva at 37 degrees C. The microgap in the connection was measured by an Image Analysis software incorporated in a high resolution scanning electron microscopy. Implant connections were filled with methylene blue by using self-adjustable precision pipettes and the quantity of leakage was determined by high sensitivity spectometry. We showed that the internal connection has lower microgaps compared to the external ones and these microgaps increased with the number of mechanical cycles. The leakage of methylene blue was higher when the external connection was performed. Microgaps and the influence of the mechanical loads are very important for the long-term behavior avoiding the bacteria colonization in the dental implants. These aspects should be known by the implantologists.
Tipus de document
Article
Versió del document
Versió acceptada
Llengua
Anglès
Matèries (CDU)
616.3 - Patologia de l'aparell digestiu. Odontologia
Paraules clau
Pàgines
7
Publicat per
Springer
Col·lecció
26; 7
Publicat a
Journal of Materials Science: Materials in Medicine
Citació recomanada
Martin Gili, David; Molmeneu Trias, Meritxell; Fernández Yagüe, Marc Antoni [et al.]. Determination of fluid leakages in the different screw-retained implant-abutment connections in a mechanical artificial mouth. Journal of Materials Science. Materials in Medicine, 2015, 26(7), [p. 1-7]. Disponible en: <https://link.springer.com/article/10.1007/s10856-015-5544-x>. Fecha de acceso: 5 may 2020. DOI: 10.1007/s10856-015-5544-x
Número de l'acord de la subvenció
info:eu-repo/grantAgreement/MINECO/6PN/MAT2012-30706
Nota
This work was supported by CICYT Spanish project MAT-2012-30706 and the Andorra and Catalan Government for CTP grants. The authors are grateful to Soadco® for your support.
Aquest element apareix en la col·lecció o col·leccions següent(s)
- Ciències de la Salut [973]
Drets
© 2020 Springer Nature Switzerland AG.