Do titanium mini-implants have the same quality of finishing and degree of contamination before and after different manipulations? An in vitro study
Autor/a
Data de publicació
2021ISSN
2075-4701
Resum
Evaluate the quality of finishing and degree of contamination before and after handling and surface treatment of titanium (Ti) orthodontic mini-implants (OMIs). A scanning electron microscope (SEM) study on ninety-six titanium OMIs was done. Energy-Dispersive X-ray Analysis (EDX) identified the present particles on manufactured OMIs surfaces. Then, OMIs were manipulated with gauze (dry sterile, soaked in chlorhexidine) and gloves (latex, nitrile) to evaluate the contamination of these handling materials. Finally, OMIs underwent surface treatments and were placed in bone to observe the contaminants they released. Roughness (Ra) and wettability with contact angle parameter (CA) were measured on these treated OMIs. Machined OMIs presented surface irregularities and were contaminated with manufacturing-process particles (carbon, plastic Polyvinyl Chloride PVC, aluminum). Hand-manipulated OMIs were also contaminated by the handling materials. OMIs surface characteristics were as follows: acid-etched (Ra ≈ 1.3 μm, CA ≈ 66°), machined (Ra ≈ 0.3 μm, CA ≈ 68°), SB (Ra ≈ 3.3 μm, CA ≈ 78°), and SBAO (Ra ≈ 3.1 μm, CA ≈ 92°). Bone was contaminated by OMIs surface defects and extra particles. Manufactured OMIs have surface contaminants that increase with clinical handling. Surface treatments (SBAO, a combination of sandblasting and anodic oxidation) increase the roughness and contact angle, which play an important role in osseointegration. Surface-treated OMIs leave titanium particles in the bone during their insertion-removal. The use of a gauze soaked in chlorhexidine is recommended when handling OMIs. Further investigations would be interesting to study more variables and confirm the present results.
Tipus de document
Article
Versió del document
Versió publicada
Llengua
Anglès
Matèries (CDU)
616 - Patologia. Medicina clínica. Oncologia
Paraules clau
Pàgines
11
Publicat per
MDPI
Col·lecció
11; 2
Publicat a
Metals
Citació recomanada
Zogheib, Tatiana; Walter-Solana, André; De La Iglesia, Fernando [et al.]. Do titanium mini-implants have the same quality of finishing and degree of contamination before and after different manipulations? An in vitro study. Metals, 2021, 11, 245. Disponible en: <https://www.mdpi.com/2075-4701/11/2/245>. Fecha de acceso: 25 feb. 2021. DOI: 10.3390/met11020245
Número de l'acord de la subvenció
info:eu-repo/grantAgreement/ES/MICINN/RTI2018-098075-B-C22
Nota
This research was funded by Spanish Government and and European Union FEDER by the concession of the grant RTI2018-098075-B-C22 and the research group Generalitat de Catalunya 2017SGR708.
Aquest element apareix en la col·lecció o col·leccions següent(s)
- Odontologia [341]
Drets
© 2021 by the authors. Licensee MDPI, Basel, Switzerland.This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Excepte que s'indiqui una altra cosa, la llicència de l'ítem es descriu com https://creativecommons.org/licenses/by/4.0/