In-vitro development of a temporal abutment screw to protect osseointegration in immediate loaded implants
Autor/a
García-Roncero, Herminio
Caballé-Serrano, Jordi
Cano Batalla, Jordi
Cabratosa-Termes, Josep
Figueras, Oscar
Data de publicació
2015-04-23ISSN
2005-7814
Resum
Purpose: In this study, a temporal abutment fixation screw, designed to fracture in a controlled way upon application of an occlusal force sufficient to produce critical micromotion was developed. The purpose of the screw was to protect the osseointegration of immediate loaded single implants. Materials and methods: Seven different screw prototypes were examined by fixing titanium abutments to 112 Mozo-Grau external hexagon implants (MG Osseous®; Mozo-Grau, S.A., Valladolid, Spain). Fracture strength was tested at 30° in two subgroups per screw: one under dynamic loading and the other without prior dynamic loading. Dynamic loading was performed in a single-axis chewing simulator using 150,000 load cycles at 50 N. After normal distribution of obtained data was verified by Kolmogorov-Smirnov test, fracture resistance between samples submitted and not submitted to dynamic loading was compared by the use of Student's t-test. Comparison of fracture resistance among different screw designs was performed by the use of one-way analysis of variance. Confidence interval was set at 95%. Results: Fractures occurred in all screws, allowing easy retrieval. Screw Prototypes 2, 5 and 6 failed during dynamic loading and exhibited statistically significant differences from the other prototypes. Conclusion: Prototypes 2, 5 and 6 may offer a useful protective mechanism during occlusal overload in immediate loaded implants.
Tipus de document
Article
Versió del document
Versió acceptada
Llengua
English
Matèries (CDU)
616.3 - Patologia de l'aparell digestiu. Odontologia
Paraules clau
Implants dentals
Odontologia -- Aparells i instruments
Implantes dentales
Odontología -- Material y equipo
Dental implants
Dentistry -- Apparatus and instruments
Pàgines
6
Publicat per
The Korean Academy of Prosthodontics
Col·lecció
7; 2
Publicat a
The Journal of Advanced Prosthodontics
Citació
García-Roncero, Herminio; Caballé-Serrano, Jordi; Cano-Batalla, Jordi [et al.]. In-vitro development of a temporal abutment screw to protect osseointegration in immediate loaded implants. The Journal of Advanced Prosthodontics, 2015 Apr. 23, 7(2), p. 160-165. Disponible en: <https://synapse.koreamed.org/DOIx.php?id=10.4047/jap.2015.7.2.160>. Fecha de acceso: 2 may. 2020. DOI: 10.4047/jap.2015.7.2.160
Aquest element apareix en la col·lecció o col·leccions següent(s)
- Odontologia [245]
Drets
© 2015 The Korean Academy of Prosthodontics. This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.