Residual stress homogenization of hybrid implants
Autor/a
Fecha de publicación
2024ISSN
2306-5354
Resumen
Objectives: Hybrid implants commonly exhibit decreased corrosion resistance and fatigue due to differences in compressive residual stresses between the smooth and rough surfaces. The main objective of this study was to investigate the influence of an annealing heat treatment to reduce the residual stresses in hybrid implants. Methodology: Commercially pure titanium (CpTi) bars were heat-treated at 800 °C and different annealing times. Optical microscopy was used to analyze the resulting grain growth kinetics. Diffractometry was used to measure residual stress after heat treatment, corrosion resistance by open circuit potential (EOCP), corrosion potentials (ECORR), and corrosion currents (ICORR) of heat-treated samples, as well as fatigue behavior by creep testing. The von Mises distribution and the resulting microstrains in heat-treated hybrid implants and in cortical and trabecular bone were assessed by finite element analysis. The results of treated hybrid implants were compared to those of untreated hybrid implants and hybrid implants with a rough surface (shot-blasted). Results: The proposed heat treatment (800 °C for 30 min, followed by quenching in water at 20 °C) could successfully homogenize the residual stress difference between the two surfaces of the hybrid implant (−20.2 MPa). It provides better fatigue behavior and corrosion resistance (p ˂ 0.05, ANOVA). Stress distribution was significantly improved in the trabecular bone. Heat-treated hybrid implants performed worse than implants with a rough surface. Clinical significance: Annealing heat treatment can be used to improve the mechanical properties and corrosion resistance of hybrid surface implants by homogenizing residual stresses.
Tipo de documento
Artículo
Versión del documento
Versión publicada
Lengua
Inglés
Materias (CDU)
616.3 - Patología del aparato digestivo. Odontología
Palabras clave
Páginas
19
Publicado por
MDPI
Colección
11; 11
Publicado en
Bioengineering
Citación
Sanjuán Álvarez, Marta; Robles Cantero, Daniel; Gil, FJ [et al.]. Residual stress homogenization of hybrid implants. Bioengineering, 2024, 11(11), 1149. Disponible en: <https://www.mdpi.com/2306-5354/11/11/1149>. Fecha de acceso: 3 mar. 2025. DOI: 10.3390/bioengineering11111149
Este ítem aparece en la(s) siguiente(s) colección(ones)
- Odontologia [334]
Derechos
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Excepto si se señala otra cosa, la licencia del ítem se describe como https://creativecommons.org/licenses/by/4.0/