Residual stress homogenization of hybrid implants
Autor/a
Data de publicació
2024ISSN
2306-5354
Resum
Objectives: Hybrid implants commonly exhibit decreased corrosion resistance and fatigue due to differences in compressive residual stresses between the smooth and rough surfaces. The main objective of this study was to investigate the influence of an annealing heat treatment to reduce the residual stresses in hybrid implants. Methodology: Commercially pure titanium (CpTi) bars were heat-treated at 800 °C and different annealing times. Optical microscopy was used to analyze the resulting grain growth kinetics. Diffractometry was used to measure residual stress after heat treatment, corrosion resistance by open circuit potential (EOCP), corrosion potentials (ECORR), and corrosion currents (ICORR) of heat-treated samples, as well as fatigue behavior by creep testing. The von Mises distribution and the resulting microstrains in heat-treated hybrid implants and in cortical and trabecular bone were assessed by finite element analysis. The results of treated hybrid implants were compared to those of untreated hybrid implants and hybrid implants with a rough surface (shot-blasted). Results: The proposed heat treatment (800 °C for 30 min, followed by quenching in water at 20 °C) could successfully homogenize the residual stress difference between the two surfaces of the hybrid implant (−20.2 MPa). It provides better fatigue behavior and corrosion resistance (p ˂ 0.05, ANOVA). Stress distribution was significantly improved in the trabecular bone. Heat-treated hybrid implants performed worse than implants with a rough surface. Clinical significance: Annealing heat treatment can be used to improve the mechanical properties and corrosion resistance of hybrid surface implants by homogenizing residual stresses.
Tipus de document
Article
Versió del document
Versió publicada
Llengua
Anglès
Matèries (CDU)
616.3 - Patologia de l'aparell digestiu. Odontologia
Paraules clau
Pàgines
19
Publicat per
MDPI
Col·lecció
11; 11
Publicat a
Bioengineering
Citació
Sanjuán Álvarez, Marta; Robles Cantero, Daniel; Gil, FJ [et al.]. Residual stress homogenization of hybrid implants. Bioengineering, 2024, 11(11), 1149. Disponible en: <https://www.mdpi.com/2306-5354/11/11/1149>. Fecha de acceso: 3 mar. 2025. DOI: 10.3390/bioengineering11111149
Aquest element apareix en la col·lecció o col·leccions següent(s)
- Odontologia [334]
Drets
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Excepte que s'indiqui una altra cosa, la llicència de l'ítem es descriu com https://creativecommons.org/licenses/by/4.0/